Overview.

BD is committed to offering our customers high-quality diagnostic and research products and takes any potential issues with product quality very seriously. Because of this commitment, we are conducting a corrective field action for certain BD flow cytometers.

We identified instruments distributed globally that may show excessive abort counts that affect the ratio of cell population events to bead events. This may result in an erroneously high absolute count. Excessive abort counts may also affect laboratory-developed tests or experiments that determine the absolute count of cell populations using BD Trucount™ beads. We are initiating  this field action so that instruments with this issue can be rapidly identified and repaired. 

We are committed to supporting our customers throughout this process.

Please visit the next page to input your information and identify whether your instruments are affected. 